Description

The direct contact between the load and the heatable and removable thermoshelves in the chamber of the Memmert vacuum oven VO ensures rapid and uniform temperature control of food, cosmetics, watches, books, PCBs or injection moulds, without the loss of heat.

Technical Specifications

Temperature

Temperaturetemperature measured through 4-wire Pt100 sensor individually for each thermoshelf
Working temperature rangemin. 5°C above ambient up to +200°C
Setting accuracy temperature0.1 °C

 

Pressure (Vacuum)

Vacuum range5 to 1100 mbar
Pressure controlDigital electronic pressure control for a speed-controlled vacuum pump. Tubing for vacuum, air and inert gas are made of material 1.4571 (ASTM 316 Ti). Programmable, digitally controlled inlet for air.
Permitted final vacuum0.01 mbar
Maximum leckage rate0.01 bar/h
Pump controlspeed control, optimised rinsing procedures for the pump membranes as well as signal output for pump ON/OFF
ConnectionVacuum connection with small flange DN16, and gas inlet with small flange DN16

 

Control Technology

Adjustable parameterstemperature (Celsius or Fahrenheit), pressure (vacuum), programme time, time zones, summertime/wintertime
Language settingGerman, English, Spanish, French, Polish, Czech, Hungarian, Italian
ControlCOCKPITTwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
TimerDigital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAITthe process time does not start until the set temperature is reached

 

Communication

InterfaceEthernet LAN, USB
Documentationprogramme stored in case of power failure
ProgrammingAtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

 

Safety

Temperature controlmechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10°C above nominal temperature
AutoSAFETYadditionally integrated over- and undertemperature protection “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic systemintegral fault diagnostics for temperature control
Alarmvisual and acoustic

 

Heating Concept

VO direct heatingfuzzy-supported MLC (Multi-Level-Controlling) microprocessor controller adapting its performance to the volume (local temperature sensing) for each thermoshelf
Thermoshelves2 connections for thermoshelves in the rear (1st and 3rd level)

 

Standard Equipment

Works calibration certificatefor +160°C at 20 mbar pressure for each supplied thermoshelf together with the vacuum oven
Internals1 thermoshelf of aluminim, material 3.3547 (ASTM B209) with integrated large-area heating

 

Stainless Steel Interior

Materialhermetically welded stainless steel interior of extremely corrosion-resistant stainless steel, material 1.4404
Interioradditional interior mountings of stainless steel, material 1.4404 (removable for cleaning), consisting of mounting at the sides with guide bars for thermoshelves and on top (diffusor) to avoid turbulences when aerating
Volume101 l
Dimensionsw(A) x h(B) x d(C): 545 x 465 x 400 mm
Max. number of internals2
Max. loading of chamber60 kg
Max. loading per internal20 kg

 

Textured Stainless Steel Casing

Dimensionsw(D) x h(E) x d(F): 710 x 767 x 550 mm (d +38mm door handle)
Doorfull-sight glass door, inside spring-loaded, 15 mm thick glazed panel in safety glass, outside with anti-splitter screen
Housingrear zinc-plated steel

 

Electrical Data

VoltageElectrical load (maximally equipped) at 230 V, 50/60 Hz

 

Ambient Conditions

Ambient temperature+5 °C to +40 °C
Set UpThe distance between the wall and the rear of the chamber must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from the wall must not be less than 8 cm.
Humidity rhmax. 80 %, non-condensing
Overvoltage categoryII
Pollution degree2

 

Packing/Shipping Data

Transport informationThe appliances must be transported upright
Customs tariff number8419 8998
Country of originFederal Republic of Germany
WEEE-Reg.-No.DE 66812464
Dimensions approx incl. cartonw x h x d: 820 x 970 x 670 mm
Net weightapprox. 110 kg
Gross weight cartonapprox. 135 kg

 